#1: CONDITIONING OF THE PH51:
THIS MUST BE DONE PRIOR TO
CALIBRATION & USE OF THIS UNIT
Remove the cap from the end of the unit and immerse
the pH51 probe into a cup or jar with 1” to 1.5” inches
of Bottle Drinking water, or 4.01 calibration solution, or
MA9015 storage solution for a minimum of 2 to 4
hours to activate the PH51 electrode.
____________________________________________
#2: CALIBRATION OF THE PH51:
After doing the conditioning and activation above
unscrew the battery compartment cap on the top of the
meter and locate the 2 calibration screws at the bottom
of the well opposite the battery compartment. Perform
the following two-point calibration procedure.
_____________________________________________
Calibration Procedure:
With the cap off Immerse the Probe into pH7.01
(M10007) buffer solution provided.
• Allow the reading to stabilize and with the small
screwdriver provided inside the case, turn the ( “7” )
calibration screw located at the bottom, inside the
battery compartment, until the LED display shows
"7.0".
• Rinse the electrode with water and now put the probe
into pH4.01 (M10004) buffer solution provided.
• Allow the reading to stabilize and with the small
screwdriver turn the ( “4” ) calibration trimmer until
the LED display shows "4.0".
* * * YOU ARE NOW CALIBRATED * * *
#3: OPERATION OF THE PH51:
• Remove the unit from your storage container cup or jar
( Remember to always store the probe wet )
• Turn the pH51 on by pressing the ON key.
• Immerse the electrode in the solution to be tested.
• Stir gently and wait for the reading to stabilize.
• After use, turn unit off by pressing the OFF key and
rinse the end of the probe with water and then return it
to jar or cup for storage.
___________________________________________
#4: STORAGE OF THE PH51:
Leave the protective cap off and put the PH51 back in a
cup or jar that has 1” to 1.5” of either Bottle Drinking
water, or 4.01 calibration solution, or MA9015 storage
solution
____________________________________________
NEVER-> LET THE PROBE DRY OUT
Always store the probe in a cup or jar that is 1to 1.5”
inches deep in Bottle Drinking water, or 4.01 calibration
solution, or MA9015 storage
_____________________________________________
NEVER -> USE DISTILLED OR
DEIONIZED WATER FOR
STORAGE PURPOSES.