4.0 PRODUCTION CALIBRATION AND TESTING PROCEDURE:
4.1 IR Software Calibration Check using Black Body Source (Calibrator 9133 or 4181)
4.1.1 Put all units inside the control room for about 3 hours where the temperature is kept at
around 23°C --25°C and humidity is 40% +/- 10%
4.1.2 Press and hold the “ALARM” button, then apply power. Press the “MEASURE” button 4
times. After 5 seconds, the unit will enter into IR calibration mode.
4.1.3 Press the “ALARM” button to skip 25°C and move to 0°C calibration point.
4.1.4 Put the unit on the test fixture. Wait for 3-4 seconds and then press the “MEASURE”
button to confirm. “PASS” will display
4.1.5 The next calibration point 50°C will appear. Repeat step3 to complete all calibration
points (0°C, 50°C, 100°C, 250°C,450°C and -20°C) according to the below calibration
table.
Blackbody [°C] Distance [inch] Stable Time (sec) Fluke
0 3 5 9133
50 2 5 9133
100 10 5 4181
250 10 3 4181
450 12 3 4181
-20 3 5 9133